User Manual 1178.8520.02 ─ 01
The "Single Sweep" and "Continuous Sweep"
softkeys
control the sweep mode for the
currently selected channel only; the sweep mode only has an effect the next time the
Sequencer activates that channel, and only for a channel-defined sequence. In this
case, a channel in single sweep mode is swept only once by the Sequencer. A channel
in continuous sweep mode is swept repeatedly.
6.4.2
Sequencer Settings
The "Sequencer" menu is available from the toolbar.
Sequencer State
Activates or deactivates the Sequencer. If activated, sequential operation according to
the selected Sequencer mode is started immediately.
Remote command:
Sequencer Mode
Defines how often which measurements are performed. The currently selected mode
softkey is highlighted blue. During an active Sequencer process, the selected mode
softkey is highlighted orange.
"Single Sequence"
Each measurement is performed once, until all measurements in all
active channels have been performed.
"Continuous Sequence"
The measurements in each active channel are performed one after
the other, repeatedly, in the same order, until sequential operation is
stopped.
This is the default Sequencer mode.
"Channel Defined Sequence"
First, a single sequence is performed. Then, only channels in continu-
ous sweep mode are repeated.
Remote command:
6.4.3
How to Set Up the Sequencer
In order to perform the configured measurements consecutively, a Sequencer function
is provided.
1. Configure a channel for each measurement configuration as required, including the
sweep mode.
Running a Sequence of Measurements