Measurements and Results
R&S
®
FSVA3000/ R&S
®
FSV3000
169
User Manual 1178.8520.02 ─ 01
●
Weighting Filter
Predefined standards can be selected via the "CP / ACLR Standard" softkey in the "Ch
Power" menu or in the "General Settings" tab of the "CP / ACLR Config" dialog box.
For details on the available standards, see
Chapter 7.3.9, "Reference: Predefined CP/
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:PRESet
User Standards ← Standard
Access
: "CP / ACLR Config" > "General Settings" tab > "Manage User Standards"
Note:
User standards created on an analyzer of the R&S FSP family are compatible to
the R&S
FSV/A. User standards created on an R&S
FSV/A, however, are not necessa-
rily compatible to the analyzers of the R&S FSP family and may not work there.
The following parameter definitions are saved in a user-defined standard:
●
Number of adjacent channels
●
Channel bandwidth of transmission (Tx), adjacent (Adj) and alternate (Alt) chan-
nels
●
Channel spacings
●
Weighting filters
●
Resolution bandwidth
●
Video bandwidth
●
Detector
●
ACLR limits and their state
●
"Sweep Time" and "Sweep Time" coupling
●
Trace and power mode
Save the current measurement settings as a user-defined standard, load a stored mea-
surement configuration, or delete an existing configuration file.
For details see
Chapter 7.3.6.4, "How to Manage User-Defined Configurations"
Remote command:
To query all available standards:
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:STANdard:CATalog?
on page 753
To load a standard:
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:PRESet
To save a standard:
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:STANdard:SAVE
on page 753
To delete a standard:
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:STANdard:DELete
Number of Sub Blocks
Defines the number of sub blocks, i.e. groups of transmission channels in an MSR sig-
nal.
For more information, see
Chapter 7.3.3.4, "Measurement on Multi-Standard Radio
Channel Power and Adjacent-Channel Power (ACLR) Measurement