Measurements and Results
R&S
®
FSVA3000/ R&S
®
FSV3000
188
User Manual 1178.8520.02 ─ 01
Figure 7-30: Adjacent channel power measurement on a CDMA2000 signal
7.3.7.2
Measurement Example 2 – Measuring Adjacent Channel Power of a W-CDMA
Uplink Signal
Test setup:
Signal
Generator
R&S FSW
Signal generator settings (e.g. R&S SMW):
Frequency:
1950 MHz
Level:
4 dBm
Modulation:
3GPP W-CDMA Reverse Link
Procedure:
1. Preset the R&S
FSV/A.
2. Set the center frequency to
1950
MHz
.
3. Select the "Channel Power ACLR" measurement function from the "Select Mea-
surement" dialog box.
4. Set the "W-CDMA 3GPP REV" standard for adjacent channel power measurement
in the "ACLR Setup" dialog box.
Channel Power and Adjacent-Channel Power (ACLR) Measurement