Measurements and Results
R&S
®
FSVA3000/ R&S
®
FSV3000
226
User Manual 1178.8520.02 ─ 01
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:STARt:
FUNCtion
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:STOP:FUNCtion
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:STARt:ABS
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:STOP:ABS
Multi-Limit Calc <n>
Defines the function used to calculate the limit line for the n-th power class for overlap-
ping ranges in Multi-SEM measurements. For details, see
"Limit calculation for individ-
"NONE"
(reference ranges only:) the limit of the reference range is used
"SUM"
Sum of the two limit lines (calculated for linear powers) is used
"MAX"
Maximum of the two limit lines is used
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<ri>:MLCalc
Min Sweep Points
Defines the minimum number of sweep points for the range.
If necessary to fulfill all minimum sweep point requirements in all ranges, the global
setting is increased. By default, each range is supplied with a minimum
of one sweep point.
For details, see
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<ri>:POINts:MINinum[:VALue]
Insert before Range / Insert after Range
Inserts a new range to the left (before) or to the right (after) of the range in which the
cursor is currently displayed. The range numbers of the currently focused range and all
higher ranges are increased accordingly. The maximum number of ranges is 30.
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<ri>:INSert
Delete Range
Deletes the currently focused range, if possible. (The reference range cannot be
deleted. A minimum of three ranges is required.) The range numbers are updated
accordingly.
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<ri>:DELete
Symmetrical Setup
Any changes to the range settings in active "Symmetrical Setup" mode lead to sym-
metrical changes in the other ranges (where possible). In particular, this means:
Spectrum Emission Mask (SEM) Measurement