Measurements and Results
R&S
®
FSVA3000/ R&S
®
FSV3000
235
User Manual 1178.8520.02 ─ 01
Selecting Storage Location - Drive/ Path/ Files
...........................................................235
.............................................................................................. 236
Selecting Storage Location - Drive/ Path/ Files
Select the storage location of the file on the instrument or an external drive.
The default storage location for the SEM settings files is:
C:\Program Files\Rohde-Schwarz\FSV3000\<version>\sem_std
.
Note:
Saving instrument settings in secure user mode.
In secure user mode, settings that are stored on the instrument are stored to volatile
memory, which is restricted to 256
MB. Thus, a "memory limit reached" error can occur
although the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
For details, see
Chapter 5.1.7, "Protecting Data Using the Secure User Mode"
Remote command:
File Name
Contains the name of the data file without the path or extension.
By default, the name of a user file consists of a base name followed by an underscore.
Multiple files with the same base name are extended by three numbers, e.g.
limit_lines_005
.
File names must be compatible with the Windows conventions for file names. In partic-
ular, they must not contain special characters such as ":", "*", "?".
For details on the filename and location, see
Chapter 11.3.2.2, "Storage Location and
Load Standard
Loads the selected measurement settings file.
Remote command:
[SENSe:]ESPectrum<sb>:PRESet[:STANdard]
Save Standard
Saves the current measurement settings for a specific standard as a file with the
defined name.
Remote command:
[SENSe:]ESPectrum<sb>:PRESet:STORe
Delete Standard
Deletes the selected standard. Standards predefined by Rohde & Schwarz can also be
deleted. A confirmation query is displayed to avoid unintentional deletion of the stan-
dard.
Spectrum Emission Mask (SEM) Measurement