Measurements and Results
R&S
®
FSVA3000/ R&S
®
FSV3000
272
User Manual 1178.8520.02 ─ 01
Remote command:
[SENSe:]SWEep:EGATe:TRACe<t>:COMMent
Period
Length of the period to be traced. The period is the same for all traces. If you change
the period for one trace, it is automatically changed for all traces.
Make sure the defined period is not longer than the total measurement time of the cur-
rent measurement. Keep in mind that the measurement time depends on the band-
width and the number of samples (see
rent measurement time is indicated as "Meas Time" in the channel bar.
Remote command:
[SENSe:]SWEep:EGATe:TRACe<t>:PERiod
Range <x> Use
Activates tracing of the defined range during a gated measurement.
Remote command:
[SENSe:]SWEep:EGATe:TRACe<t>[:STATe<gr>]
Range <x> Start/Stop
Defines the start and stop points of the range within the tracing period. Make sure the
value for the stopping time is smaller than the length of the period.
Note:
You can define the time values with a greater numerical resolution than is dis-
played; the values are only rounded for display.
Remote command:
[SENSe:]SWEep:EGATe:TRACe<t>:STARt<gr>
[SENSe:]SWEep:EGATe:TRACe<t>:STOP<gr>
7.8.5.3
Scaling for Statistics Diagrams
Access
: "Overview" > "Amplitude" > "Scale" tab
Or
: [AMPT] > "Scale Config"
For statistics displays, scale settings are available for both the y-axis and the x-axis.
The remote commands required to perform these tasks are described in
ter 14.5.8.4, "Scaling the Diagram"
Statistical Measurements (APD, CCDF)