Measurements and Results
R&S
®
FSVA3000/ R&S
®
FSV3000
298
User Manual 1178.8520.02 ─ 01
Auto measurement
Use the
function to determine the required measurement parameters automat-
ically.
The following settings are determined:
●
Peak markers
●
Center frequency of 2 narrowband signals creating intermodulation products
See also
Chapter 7.1, "Auto Measurements"
The remote commands required to perform these tasks are described in
ter 14.5.11, "Measuring the Third Order Intercept Point"
Marker 1 / Marker 2 / Marker 3 / Marker 4
................................................................. 298
Marker 1 / Marker 2 / Marker 3 / Marker 4
Indicates the detected characteristic values as determined by the TOI measurement
(see
The marker positions can be edited; the TOI is then recalculated according to the new
marker values.
To reset all marker positions automatically, use the
function.
Remote command:
CALCulate<n>:DELTamarker<m>:X:RELative?
Search Signals
Performs a new search on the input signals and recalculates the TOI according to the
measured values.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:TOI:SEARchsignal ONCE
7.11.5
How to Determine the Third Order Intercept
The precise TOI for the R&S
FSV/A in relation to the input signals is provided in the
data sheet.
For remote operation, see
Chapter 14.5.11.2, "Programming Example: Measuring the
1. Apply a two-tone signal with equal carrier levels to the R&S
FSV/A input.
2. On the R&S
FSV/A, press the [MEAS] key.
3. Select the "Third Order Intercept" measurement function from the "Select Measure-
ment" dialog box.
Third Order Intercept (TOI) Measurement