Common Measurement Settings
R&S
®
FSVA3000/ R&S
®
FSV3000
337
User Manual 1178.8520.02 ─ 01
For more information on configuring interfaces, see
Chapter 13.1.1, "Remote Control
GPIB Address / TCPIP Address / Computer Name
................................................... 337
...........................................................................................338
................................................................................338
................................................................................................ 338
Generator Type
Selects the generator type and thus defines the generator setup file to use.
For an overview of supported generators, see
"Overview of Supported Generators"
on page 328. For information on generator setup files, see
Remote command:
SYSTem:COMMunicate:RDEVice:GENerator<gen>:TYPE
Interface
Type of interface connection used. The following interfaces are currently supported:
●
GPIB
●
TCP/IP (not by all generators)
For details on which signal generators support which interfaces, see the documenta-
tion of the corresponding signal generator.
Remote command:
SYSTem:COMMunicate:RDEVice:GENerator<gen>:INTerface
TTL Handshake
If available for the specified generator type, this option activates TTL synchronization
via handshake for GPIB connections.
Using the TTL interface allows for considerably higher measurement rates than pure
GPIB control, because the frequency stepping of the R&S
FSV/A is directly coupled
with the frequency stepping of the generator.
For more information on TTL synchronization, see
For an overview of which generators support TTL synchronization see
Remote command:
SYSTem:COMMunicate:RDEVice:GENerator<gen>:LINK
GPIB Address / TCPIP Address / Computer Name
For LAN connections: TCP/IP address of the signal generator
For GPIB connections: GPIB address of the signal generator.
Data Input and Output