Common Measurement Settings
R&S
®
FSVA3000/ R&S
®
FSV3000
369
User Manual 1178.8520.02 ─ 01
scaling is useful when large ranges of values must be combined in one display. Loga-
rithmic scaling is best applied to measurement values in logarithmic units (dB, dBm
etc.).
In addition to linear or logarithmic scaling, the vertical axis can be set to display either
absolute or relative values. Absolute values show the measured levels, while relative
values show the difference between the measured level and the defined reference
level. Relative values are indicated in percent for linear scaling, and in dB for logarith-
mic scaling.
8.4.2
Amplitude Settings
Access
: "Overview" > "Amplitude"
Amplitude settings determine how the R&S
FSV/A must process or display the expec-
ted input power levels.
The remote commands required to define these settings are described in
ter 14.7.3.1, "Amplitude Settings"
Shifting the Display ( Offset )
........................................................................ 370
Setting the Reference Level Automatically ( Auto Level )
.............................371
.............................................................................371
...................................................................................... 371
Amplitude and Vertical Axis Configuration