Common Analysis and Display Functions
R&S
®
FSVA3000/ R&S
®
FSV3000
509
User Manual 1178.8520.02 ─ 01
Remote command:
CALCulate<n>:SPECtrogram:THReedim[:STATe]
Select Frame
Selects a specific frame, loads the corresponding trace from the memory, and displays
it in the Spectrum window.
Note that activating a marker or changing the position of the active marker automati-
cally selects the frame that belongs to that marker.
This function is only available in single sweep mode or if the sweep is stopped, and
only if a spectrogram is selected.
The most recent frame is number 0, all previous frames have a negative number.
Remote command:
CALCulate<n>:SPECtrogram:FRAMe:SELect
History Depth
Sets the number of frames that the R&S
FSV/A stores in its memory.
The maximum number of frames depends on the
.
If the memory is full, the R&S
FSV/A deletes the oldest frames stored in the memory
and replaces them with the new data.
Remote command:
CALCulate<n>:SPECtrogram:HDEPth
3-D Display Depth
Defines the number of frames displayed in a 3-dimensional spectrogram.
For details see
"Three-Dimensional Spectrograms"
Time Stamp
Activates and deactivates the timestamp. The timestamp shows the system time while
the measurement is running. In single sweep mode or if the sweep is stopped, the
timestamp shows the time and date of the end of the sweep.
When active, the timestamp replaces the display of the frame number.
Remote command:
CALCulate<n>:SPECtrogram:TSTamp[:STATe]
CALCulate<n>:SPECtrogram:TSTamp:DATA?
Color Mapping
Opens the "Color Mapping" dialog.
For details see
Continuous Sweep / Run Cont
After triggering, starts the sweep and repeats it continuously until stopped. This is the
default setting.
After triggering, starts the measurement and repeats it continuously until stopped.
Trace Configuration