The following units are available:
●
dB
●
dBm
●
dBmV
●
dBμV
●
dBμV/m
●
dBμA
●
dBμA/m
●
dBpW
●
dBpT
Compatibility
Indicates whether the transducer factors are compatible with the current measurement
settings.
For more information on which conditions a transducer line must fulfill to be compati-
ble, see
Chapter 12.5.1, "Basics on Transducer Factors"
Activating / Deactivating
Activates/deactivates the transducer line. Up to 8 transducer lines can be active at the
same time.
Remote command:
[SENSe:]CORRection:TRANsducer:SELect
[SENSe:]CORRection:TRANsducer[:STATe]
Comment
An optional description of the transducer line.
Included Transducer Lines in Overview (View Filter)
Defines which of the stored transducer lines are included in the overview. The view can
be restricted to compatible transducer lines only or include all transducer lines found.
Whether a line is compatible or not is indicated in the
setting.
Adjust Ref Level
Activates or deactivates the automatic adjustment of the reference level to the selected
transducer factor.
"Auto"
Activates the automatic adjustment. The original dynamic range is
restored by shifting the reference level by the maximum transducer
factor.
"Manual"
Deactivates the automatic adjustment. Adjust the reference level via
the "Amplitude" menu.
Remote command:
[SENSe:]CORRection:TRANsducer:ADJust:RLEVel[:STATe]
Create New Line
Create a new transducer line.
Remote command:
[SENSe:]CORRection:TRANsducer:SELect
Transducers