Figure 5-13: "MultiView" tab with active Sequencer
, the "Spectrum 2" measurement is currently active (indicated by the
"channel active" icon in the tab label).
3. Stop the Sequencer by tapping the "Sequencer" softkey again.
5.3.5
Setting and Moving a Marker
Markers are useful to determine the position of particular effects in the trace. The most
common use is to determine a peak, which is the default setting when you activate a
marker. We will set a marker on the peak in our first Spectrum measurement.
1. In the "MultiView" tab, double-tap the "Spectrum" window (frequency sweep with
spectrogram display) to return to the "Spectrum" channel.
2. Tap the spectrum display to set the focus on that window.
3.
Press the "Split/Maximize" key on the front panel to maximize the spectrum win-
dow, as we currently do not need the spectrogram display.
4. Press the "RUN SINGLE" key on the front panel to perform a single sweep so we
have a fixed trace to set a marker on.
5. Press the [MKR] key on the front panel to display the "Marker" menu.
Marker 1 is activated and automatically set to the maximum of trace 1. The marker
position and value is indicated in the diagram area as M1[1].
Trying Out the Instrument