Network and Remote Operation
R&S
®
FSVA3000/ R&S
®
FSV3000
723
User Manual 1178.8520.02 ─ 01
13.6.7.1
How to Configure the R&S
FSV/A for Remote Operation via Remote Desktop
1. Create a fixed IP address for the TCP/IP protocol as described in
"How to Assign the IP Address"
Note:
To avoid problems, use a fixed IP address.
When a DHCP server is used, a new IP address is assigned each time the instru-
ment is restarted. This address must first be determined on the instrument itself.
Thus, using a DHCP server is not suitable for remote operation of the R&S
FSV/A
via Remote Desktop.
2.
Select the "Windows" icon in the toolbar to access the operating system.
3. In the Windows "Start" menu, select "Settings > System".
4. Search for "remote access".
5. Select "Allow remote access to your computer".
6. Define which users are to be given access to the R&S
FSV/A via Remote Desktop.
Note
: The user account under which configuration is carried out is automatically
enabled for Remote Desktop.
a) Select the "Select Users" button.
b) Select the users or create new user accounts as described in
How to Set Up a Network and Remote Control