Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
735
User Manual 1178.8520.02 ─ 01
Parameters:
<Contents>
Contents of the service request enable register in decimal form.
Bit 6 (MSS mask bit) is always 0.
Range:
0 to 255
*STB?
Status byte query
Reads the contents of the status byte in decimal form.
Usage:
Query only
*TRG
Trigger
Triggers all actions waiting for a trigger event. In particular, *TRG generates a manual
trigger signal. This common command complements the commands of the
TRIGger
subsystem.
*TRG
corresponds to the
INITiate:IMMediate
Usage:
Event
*TST?
Self-test query
Initiates self-tests of the instrument and returns an error code.
Note:
If you start a self-test remotely, then select the "Local" softkey while the test is
still running, the instrument only returns to the manual operation state after the test is
completed. In this case, the self-test cannot be aborted.
Return values:
<ErrorCode>
integer > 0 (in decimal format)
An error occurred.
(For details, see the Service Manual supplied with the instru-
ment).
0
No errors occurred.
Usage:
Query only
*WAI
Wait to continue
Common Commands