Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
736
User Manual 1178.8520.02 ─ 01
Prevents servicing of the subsequent commands until all preceding commands have
been executed and all signals have settled (see also command synchronization and
).
Usage:
Event
14.4
Selecting the Operating Mode and Application
The following commands are required to select the operating mode or the application
and to configure a Sequencer in a remote environment.
Selecting the Mode and Applications
....................................................................736
Performing a Sequence of Measurements
............................................................740
Programming Example: Performing a Sequence of Measurements
14.4.1
Selecting the Mode and Applications
...................................................................................... 736
............................................................................................ 737
........................................................................................ 737
DISPlay:ATAB
<State>
This command switches between the MultiView tab and the most recently displayed
channel. If only one channel is active, this command has no effect.
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
INSTrument:CREate:DUPLicate
This command duplicates the currently selected channel, i.e creates a new channel of
the same type and with the identical measurement settings. The name of the new
channel is the same as the copied channel, extended by a consecutive number (e.g.
"IQAnalyzer" -> "IQAnalyzer 2").
The channel to be duplicated must be selected first using the
INST:SEL
command.
Selecting the Operating Mode and Application