Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
741
User Manual 1178.8520.02 ─ 01
Manual operation:
See
INITiate:SEQuencer:IMMediate
This command starts a new sequence of measurements by the Sequencer.
Before this command can be executed, the Sequencer must be activated (see
Example:
SYST:SEQ ON
Activates the Sequencer.
INIT:SEQ:MODE SING
Sets single sequence mode so each active measurement will be
performed once.
INIT:SEQ:IMM
Starts the sequential measurements.
Manual operation:
See
INITiate:SEQuencer:MODE
<Mode>
This command selects the way the R&S
FSV/A application performs measurements
sequentially.
Before this command can be executed, the Sequencer must be activated (see
A detailed programming example is provided in
Chapter 14.4.3, "Programming Exam-
ple: Performing a Sequence of Measurements"
Note:
In order to synchronize to the end of a sequential measurement using *OPC,
*OPC? or *WAI you must use
SINGle
Sequence mode.
For details on synchronization see
Chapter 13.1.6, "Command Sequence and Syn-
Parameters:
<Mode>
SINGle
Each measurement is performed once (regardless of the chan-
nel's sweep mode), considering each channels' sweep count,
until all measurements in all active channels have been per-
formed.
CONTinuous
The measurements in each active channel are performed one
after the other, repeatedly (regardless of the channel's sweep
mode), in the same order, until the Sequencer is stopped.
CDEFined
First, a single sequence is performed. Then, only those channels
in continuous sweep mode (
INIT:CONT ON
) are repeated.
*RST:
CONTinuous