Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
747
User Manual 1178.8520.02 ─ 01
ON | 1
Continuous sweep
OFF | 0
Single sweep
*RST:
1
<State>
ON | OFF | 0 | 1
ON | 1
Continuous sweep
OFF | 0
Single sweep
*RST:
0
Example:
INIT:CONT OFF
Switches the sweep mode to single sweep.
INIT:CONT ON
Switches the sweep mode to continuous sweep.
Manual operation:
See
See
See
" Continuous Sweep / Run Cont "
INITiate<n>[:IMMediate]
This command starts a (single) new measurement.
With sweep count or average count > 0, this means a restart of the corresponding
number of measurements. With trace mode MAXHold, MINHold and AVERage, the
previous results are reset on restarting the measurement.
You can synchronize to the end of the measurement with *OPC, *OPC? or *WAI.
For details on synchronization see
Chapter 13.1.6, "Command Sequence and Syn-
Suffix:
<n>
.
irrelevant
Example:
INIT:CONT OFF
Switches to single sweep mode.
DISP:WIND:TRAC:MODE AVER
Switches on trace averaging.
SWE:COUN 20
Sets the sweep counter to 20 sweeps.
INIT;*WAI
Starts the measurement and waits for the end of the 20 sweeps.
Manual operation:
See
See
See
Configuring and Performing Measurements