User Manual 1178.8520.02 ─ 01
OBANdwidth | OBWidth
Occupied bandwidth.
Returns the occupied bandwidth in Hz.
Manual operation:
See
See
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>:SELect
<MeasType>
This command selects a power measurement and turns the measurement on.
Suffix:
<n>
.
<m>
<sb>
irrelevant
Parameters:
<MeasType>
ACPower | MCACpower
Adjacent channel leakage ratio (ACLR), also known as adjacent
channel power or multicarrier adjacent channel.
The R&S
FSV/A performs the measurement on the trace
selected with
CPOWer
Channel power measurement with a single carrier.
The R&S
FSV/A performs the measurement on the trace
selected with
OBANdwidth | OBWidth
Occupied bandwidth measurement.
The R&S
FSV/A performs the measurement on the trace that
marker 1 is positioned on.
Manual operation:
See
See
CALCulate<n>:MARKer<m>:FUNCtion:POWer<sb>[:STATe]
<State>
This command turns a power measurement on and off.
Suffix:
<n>
.
irrelevant
<m>
irrelevant
<sb>
irrelevant
Parameters:
<State>
ON | OFF | 1 | 0
ON | 1
The power measurement selected with
MARKer<m>:FUNCtion:POWer<sb>:SELect
Configuring and Performing Measurements