Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
777
User Manual 1178.8520.02 ─ 01
If you have defined an absolute limit as well as a relative limit, the R&S
FSV/A uses the
lower value for the limit check.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
<gap>
Gap (CACLR) channel number
Parameters:
<Limit>
Defines the relative limit of the specified gap channel in dB.
Default unit: DB
<UpperLimit>
Optional: Defines the relative upper limit of the specified gap
channel in dB.
Default unit: dB
Example:
CALC:LIM:ACP:GAP1:REL 3dB,0
Manual operation:
See
CALCulate<n>:LIMit<li>:ACPower:GAP<gap>[:CACLr][:RELative]:STATe
<State>
This command turns the relative limit check for the specified gap (CACLR) channel on
and off.
You have to activate the general ACLR limit check before using this command with
CALCulate<n>:LIMit<li>:ACPower[:STATe]
.
Suffix:
<n>
.
irrelevant
<li>
irrelevant
<gap>
Gap (CACLR) channel number
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:LIM:ACP:GAP1:REL:STAT ON
Manual operation:
See
[SENSe:]POWer:ACHannel:AGCHannels
<State>
This command activates or deactivates gap channels in an MSR signal.
Configuring and Performing Measurements