Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
779
User Manual 1178.8520.02 ─ 01
[SENSe:]POWer:ACHannel:FILTer[:STATe]:GAP<gap>
<State>
This command turns the weighting filter for the specified gap (CACLR) channel on and
off.
Suffix:
<gap>
.
1..n
Gap (CACLR) channel number
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Manual operation:
See
[SENSe:]POWer:ACHannel:GAP<gap>:MSIZe
<Bandwidth>
If the gap between the sub blocks does not exceed the specified bandwidth, the gap
channels are not displayed in the diagram, and the gap channel results are not calcula-
ted in the result summary.
Suffix:
<gap>
.
1..n
Gap channel number
Parameters:
<Bandwidth>
numeric value in Hz
*RST:
gap1: 5 MHz; gap2: 10 MHz
Default unit: Hz
Manual operation:
See
" Minimum gap size to show Gap 1 / Minimum gap size to
[SENSe:]POWer:ACHannel:SPACing:GAP<gap>
<Spacing>
This command defines the distance from sub block to the specified gap channel.
The channels in the upper gap are identical to those in the lower gap. Thus, in the
R&S
FSV/A MSR ACLR measurement, only 2 gap channels are configured.
The spacing for gap channels is defined in relation to the outer edges of the surround-
ing sub blocks, i.e.
Spacing = [CF of the gap channel] - [left sub block center] + ([RF bandwidth of left sub
block] /2)
(See also
and
.)
Suffix:
<gap>
.
Gap channel number
Parameters:
<Spacing>
numeric value in Hz
*RST:
2.5 MHz
Default unit: HZ
Configuring and Performing Measurements