Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
800
User Manual 1178.8520.02 ─ 01
<ri>
1..n
Selects the measurement range.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:INSert
<Mode>
This command inserts a new SEM range and updates the range numbers accordingly.
Suffix:
<sb>
.
1..n
<ri>
1..n
Selects the SEM range.
Parameters:
<Mode>
AFTer | BEFore
AFTer
Inserts a range after the selected range.
BEFore
Inserts a range before the selected range.
Manual operation:
See
" Insert before Range / Insert after Range "
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:ABSolute:STARt
<Level>
This command defines an absolute limit for a SEM range.
Unlike manual operation, you can define an absolute limit anytime and regardless of
the limit check mode.
Suffix:
<sb>
.
1..n
<ri>
1..n
Selects the measurement range.
<li>
1..n
Power class for which the limit is defined.
Parameters:
<Level>
Absolute limit at the start frequency of a SEM range.
Range:
-400 to 400
*RST:
-13
Default unit: dBm
Example:
SENSe:ESPectrum:RANGe:LIMit:ABSolute:STARt -10
Chapter 14.5.6.11, "Example: SEM
Manual operation:
See
" Abs Limit Start / Stop <n> "
Configuring and Performing Measurements