Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
808
User Manual 1178.8520.02 ─ 01
Parameters:
<Transducer>
String containing the transducer file name, including the path
information.
Manual operation:
See
[SENSe:]ESPectrum<sb>:SSETup
<State>
Enables or disables symmetrical configuration of the range settings.
See
Chapter 7.6.4.1, "Ranges and Range Settings"
Suffix:
<sb>
.
1..n
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST:
0
Manual operation:
See
14.5.6.5
Configuring the Reference Range
The following commands define the reference range for the SEM sweep list.
...................................................................................... 808
[SENSe:]ESPectrum<sb>:FILTer[:RRC]:ALPHa
.................................................................809
[SENSe:]ESPectrum<sb>:FILTer[:RRC][:STATe]
................................................................809
................................................................................809
.................................................................................... 809
[SENSe:]ESPectrum<sb>:BWID
<Bandwidth>
This command defines the channel bandwidth of the reference range.
The bandwidth is available if the power reference is the channel power.
Suffix:
<sb>
.
1..n
Sub block in a Multi-SEM measurement
Parameters:
<Bandwidth>
minimum span ≤ value ≤ span of reference range
*RST:
3.84 MHz
Default unit: Hz
Manual operation:
See
Configuring and Performing Measurements