Getting Started
R&S
®
FSVA3000/ R&S
®
FSV3000
81
User Manual 1178.8520.02 ─ 01
The following types of information are displayed, if applicable.
Table 5-5: Common settings displayed in the channel bar
"SGL"
The sweep is set to single sweep mode.
"Sweep
Count"
The current signal count for measurement tasks that involve a specific number of subse-
quent sweeps
(see "Sweep Count" setting in "Sweep settings" in the User Manual)
"TRG"
Trigger source
(for details see "Trigger settings" in the User Manual)
●
EXT
: External
●
IFP
: IF power (+trigger bandwidth)
●
PSE
: Power sensor
●
RFP
: RF power
●
SQL
: Squelch
●
TIM
: Time
●
VID
: Video
"6dB"/"RRC"/"
CHN"
Filter type for sweep bandwidth
"PA"
/
Ext "PA"
The preamplifier is activated. / Data compensation is performed using data from the
(optional) external preamplifier.
"YIG Bypass"
The YIG filter is deactivated.
"GAT"
The frequency sweep is controlled via the TRIGGER INPUT connector.
"TDF"
A transducer factor is activated.
"75
Ω
"
The input impedance of the instrument is set to 75
Ω
.
"FRQ"
A frequency offset
≠
0 Hz is set.
"DC/AC"
DC or AC coupling is used for the input.
<
"NOR"
|
"APR"
>
"Ext. Gen"
The R&S
FSV/A is controlling an external generator (requires option R&S
FSV3-B10).
NOR
: the measurements are normalized with the results of the external generator cali-
bration
APR
(approximation): the measurements are normalized with the results of the external
generator calibration; however, the measurement settings have been changed since cali-
bration
If neither label is displayed, no calibration has been performed yet or normalization is not
active.
Chapter 8.2.4, "Optional External Generator Control"
"LVL"
A level offset is applied to the external generator signal (only if external generator control
is active).
Operating the Instrument