Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
839
User Manual 1178.8520.02 ─ 01
Suffix:
<n>
.
irrelevant
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
CALC:STAT:CCDF ON
Switches on the CCDF measurement.
14.5.8.2
Configuring Statistical Measurements
The following commands configure the measurement.
Useful commands for configuring statistical measurements described elsewhere:
●
[SENSe:]BANDwidth[:RESolution]
●
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel
(Make sure the specified reference level is higher than the measured peak value,
see
Remote commands exclusive to statistical measurements:
CALCulate<n>:MARKer<m>:Y:PERCent
.......................................................................... 839
CALCulate<n>:STATistics:NSAMples
............................................................................... 839
CALCulate<n>:MARKer<m>:Y:PERCent
<Probability>
This command sets a marker to a particular probability value. You can query the corre-
sponding level with
.
Using the command turns delta markers into normal markers.
Suffix:
<n>
.
<m>
Parameters:
<Probability>
Range:
0 % to 100 %
Default unit: %
Example:
CALC1:MARK:Y:PERC 95PCT
Positions marker 1 to a probability of 95 %.
Manual operation:
See
CALCulate<n>:STATistics:NSAMples
<Samples>
This command defines the number of samples included in the analysis of statistical
measurement functions.
Suffix:
<n>
.
Configuring and Performing Measurements