Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
891
User Manual 1178.8520.02 ─ 01
CALCulate<n>:MARKer<m>:FUNCtion:STRack:THReshold
<Level>
This command defines the threshold level for the signal tracking process.
Note that you have to turn on signal tracking before you can use the command.
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Parameters:
<Level>
Range:
-130 dBm to 30 dBm
*RST:
-120 dBm
Default unit: DBM
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:STRack:TRACe
<TraceNumber>
This command selects the trace on which the largest signal is searched for.
Suffix:
<n>
.
irrelevant
<m>
irrelevant
Parameters:
<TraceNumber>
1 to 6
Range:
1 to 6
*RST:
1
Manual operation:
See
14.7.2
Configuring Bandwidth and Sweep Settings
The commands required to configure the bandwidth, sweep and filter settings in a
remote environment are described here. The tasks for manual operation are described
in
Chapter 8.5, "Bandwidth, Filter and Sweep Configuration"
Configuring the Bandwidth and Filter
....................................................................891
..........................................................................................894
14.7.2.1
Configuring the Bandwidth and Filter
.........................................................................................892
[SENSe:]BANDwidth[:RESolution]
................................................................................... 892
[SENSe:]BWIDth[:RESolution]:AUTO
............................................................................... 892
[SENSe:]BANDwidth[:RESolution]:AUTO
..........................................................................892
[SENSe:]BWIDth[:RESolution]:RATio
............................................................................... 892
[SENSe:]BANDwidth[:RESolution]:RATio
..........................................................................892
[SENSe:]BWIDth[:RESolution]:TYPE
................................................................................893
[SENSe:]BANDwidth[:RESolution]:TYPE
.......................................................................... 893
Setting Basic Measurement Parameters