Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
892
User Manual 1178.8520.02 ─ 01
............................................................................................ 893
........................................................................................ 893
...................................................................................893
........................................................................................ 894
...................................................................................894
.........................................................................................894
................................................................................... 894
[SENSe:]BWIDth[:RESolution]
<Bandwidth>
[SENSe:]BANDwidth[:RESolution]
<Bandwidth>
This command defines the resolution bandwidth and decouples the resolution band-
width from the span.
For statistics measurements, this command defines the
demodulation
bandwidth.
Parameters:
<Bandwidth>
refer to data sheet
*RST:
RBW: AUTO is set to ON; DBW: 3MHz
Default unit: Hz
Example:
BAND 1 MHz
Sets the resolution bandwidth to 1 MHz
Manual operation:
See
See
See
[SENSe:]BWIDth[:RESolution]:AUTO
<State>
[SENSe:]BANDwidth[:RESolution]:AUTO
<State>
This command couples and decouples the resolution bandwidth to the span.
Parameters:
<State>
ON | OFF | 0 | 1
*RST:
1
Example:
BAND:AUTO OFF
Switches off the coupling of the resolution bandwidth to the
span.
Manual operation:
See
See
[SENSe:]BWIDth[:RESolution]:RATio
<Ratio>
[SENSe:]BANDwidth[:RESolution]:RATio
<Ratio>
This command defines the ratio between the resolution bandwidth (Hz) and the span
(Hz).
Note that the ratio defined with this remote command (RBW/span) is reciprocal to that
of the coupling ratio (span/RBW).
Setting Basic Measurement Parameters