Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
900
User Manual 1178.8520.02 ─ 01
Suffix:
<n>
.
irrelevant
<t>
irrelevant
Parameters:
<ReferenceLevel>
The unit is variable.
Range:
see datasheet
*RST:
0 dBm
Default unit: DBM
Example:
DISP:TRAC:Y:RLEV -60dBm
Manual operation:
See
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel:OFFSet
<Offset>
This command defines a reference level offset (for all traces in all windows).
Suffix:
<n>
.
irrelevant
<t>
irrelevant
Parameters:
<Offset>
Range:
-200 dB to 200 dB
*RST:
0dB
Default unit: DB
Example:
DISP:TRAC:Y:RLEV:OFFS -10dB
Manual operation:
See
"Shifting the Display ( Offset )"
See
" Shifting the Display ( Offset )"
[SENSe:]POWer:NCORrection
<State>
This command turns noise cancellation on and off.
If noise cancellation is on, the R&S
FSV/A performs a reference measurement to
determine its inherent noise and subtracts the result from the channel power measure-
ment result (first active trace only).
For more information see
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
POW:NCOR ON
Manual operation:
See
14.7.3.2