Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
911
User Manual 1178.8520.02 ─ 01
Note that any RF attenuation or preamplification is considered when the trigger level is
analyzed. If defined, a reference level offset is also considered.
Parameters:
<TriggerLevel>
Range:
-130 dBm to 30 dBm
*RST:
-20 dBm
Default unit: DBM
Example:
TRIG:LEV:IQP -30DBM
Manual operation:
See
TRIGger[:SEQuence]:LEVel:VIDeo
<Level>
This command defines the level the video signal must exceed to cause a trigger event.
Note that any RF attenuation or preamplification is considered when the trigger level is
analyzed.
Parameters:
<Level>
Range:
0 PCT to 100 PCT
*RST:
50 PCT
Default unit: PCT
Example:
TRIG:LEV:VID 50PCT
Manual operation:
See
TRIGger[:SEQuence]:SLOPe
<Type>
For all trigger sources except time you can define whether triggering occurs when the
signal rises to the trigger level or falls down to it.
Parameters:
<Type>
POSitive | NEGative
POSitive
Triggers when the signal rises to the trigger level (rising edge).
NEGative
Triggers when the signal drops to the trigger level (falling edge).
*RST:
POSitive
Example:
TRIG:SLOP NEG
Manual operation:
See
TRIGger[:SEQuence]:SOURce
<Source>
This command selects the trigger source.
For details on trigger sources see
Using a trigger or gated measurements turns the squelch off (see
Note on external triggers: