Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
920
User Manual 1178.8520.02 ─ 01
[SENSe:]ADJust:CONFigure:DURation:MODE
<Mode>
In order to determine the ideal reference level, the R&S
FSV/A performs a measure-
ment on the current input data. This command selects the way the R&S
FSV/A deter-
mines the length of the measurement .
Parameters:
<Mode>
AUTO
The R&S
FSV/A determines the measurement length automati-
cally according to the current input data.
MANual
The R&S
FSV/A uses the measurement length defined by
[SENSe:]ADJust:CONFigure:DURation
*RST:
AUTO
Manual operation:
See
" Resetting the Automatic Measurement Time ( Meastime
See
" Changing the Automatic Measurement Time ( Meastime
[SENSe:]ADJust:CONFigure:HYSTeresis:LOWer
<Threshold>
Parameters:
<Threshold>
Range:
0 dB to 200 dB
*RST:
+1 dB
Default unit: dB
Example:
SENS:ADJ:CONF:HYST:LOW 2
For an input signal level of currently 20 dBm, the reference level
will only be adjusted when the signal level falls below 18
dBm.
Manual operation:
See
[SENSe:]ADJust:CONFigure:HYSTeresis:UPPer
<Threshold>
Parameters:
<Threshold>
Range:
0 dB to 200 dB
*RST:
+1 dB
Default unit: dB
Example:
SENS:ADJ:CONF:HYST:UPP 2
Example:
For an input signal level of currently 20 dBm, the reference level
will only be adjusted when the signal level rises above 22
dBm.
Manual operation:
See
[SENSe:]ADJust:CONFigure:TRIGger
<State>
Defines the behavior of the measurement when adjusting a setting automatically (using
SENS:ADJ:LEV ON
, for example).
Setting Basic Measurement Parameters