Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
942
User Manual 1178.8520.02 ─ 01
Example:
SYST:COMM:RDEV:PMET:COUN?
Usage:
Query only
Manual operation:
See
SYSTem:COMMunicate:RDEVice:PMETer<p>:DEFine
<Placeholder>, <Type>,
<Interface>, <SerialNo>
This command assigns the power sensor with the specified serial number to the
selected power sensor index (configuration).
The query returns the power sensor type and serial number of the sensor assigned to
the specified index.
Suffix:
<p>
.
Power sensor index
Parameters:
<Placeholder>
Currently not used
<Type>
Detected power sensor type, e.g. "NRP-Z81".
<Interface>
Interface the power sensor is connected to; always "USB"
<SerialNo>
Serial number of the power sensor assigned to the specified
index
Example:
SYST:COMM:RDEV:PMET2:DEF '','NRP-Z81','',
'123456'
Assigns the power sensor with the serial number '123456' to the
configuration "Power Sensor 2".
SYST:COMM:RDEV:PMET2:DEF?
Queries the sensor assigned to "Power Sensor 2".
Result:
'','NRP-Z81','USB','123456'
The NRP-Z81 power sensor with the serial number '123456' is
assigned to the "Power Sensor 2".
Manual operation:
See
Configuring Power Sensor Measurements
CALibration:PMETer<p>:ZERO:AUTO ONCE
................................................................... 943
CALCulate<n>:PMETer<p>:RELative[:MAGNitude]
............................................................943
CALCulate<n>:PMETer<p>:RELative[:MAGNitude]:AUTO ONCE
....................................... 943
CALCulate<n>:PMETer<p>:RELative:STATe
.....................................................................944
[SENSe:]PMETer<p>:DCYCle[:STATe]
............................................................................. 944
[SENSe:]PMETer<p>:DCYCle:VALue
...............................................................................945
................................................................................... 945
[SENSe:]PMETer<p>:FREQuency:LINK
........................................................................... 946
...........................................................................................946
[SENSe:]PMETer<p>:MTIMe:AVERage:COUNt
.................................................................946
Setting Basic Measurement Parameters