Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
949
User Manual 1178.8520.02 ─ 01
UNIT<n>:PMETer<p>:POWer:RATio
<Unit>
This command selects the unit for relative power sensor measurements.
Suffix:
<n>
.
irrelevant
<p>
Power sensor index
Parameters:
<Unit>
DB | PCT
*RST:
DB
Example:
UNIT:PMET:POW:RAT DB
Manual operation:
See
Triggering with Power Sensors
[SENSe:]PMETer<p>:TRIGger:DTIMe
..............................................................................949
[SENSe:]PMETer<p>:TRIGger:HOLDoff
........................................................................... 949
[SENSe:]PMETer<p>:TRIGger:HYSTeresis
.......................................................................950
[SENSe:]PMETer<p>:TRIGger:LEVel
............................................................................... 950
[SENSe:]PMETer<p>:TRIGger:SLOPe
............................................................................. 951
[SENSe:]PMETer<p>:TRIGger[:STATe]
.............................................................................951
[SENSe:]PMETer<p>:TRIGger:DTIMe
<Time>
This command defines the time period that the input signal has to stay below the IF
power trigger level before the measurement starts.
Suffix:
<p>
.
Power sensor index
Parameters:
<Time>
Range:
0 s to 1 s
Increment: 100 ns
*RST:
100 µs
Default unit: S
Example:
PMET2:TRIG:DTIMe 0.001
[SENSe:]PMETer<p>:TRIGger:HOLDoff
<Holdoff>
This command defines the trigger holdoff for external power triggers.
Suffix:
<p>
.
Power sensor index
Setting Basic Measurement Parameters