Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
967
User Manual 1178.8520.02 ─ 01
CALCulate<n>:SGRam:FRAMe:COUNt
<Frames>
CALCulate<n>:SPECtrogram:FRAMe:COUNt
<Frames>
This command defines the number of frames to be recorded in a single sweep.
This value applies to all spectrograms in the channel.
Suffix:
<n>
.
irrelevant
Parameters:
<Frames>
The maximum number of frames depends on the history depth.
Range:
1 to history depth
Increment: 1
*RST:
1
Example:
INIT:CONT OFF
Selects single sweep mode.
CALC:SGR:FRAM:COUN 200
Sets the number of frames to 200.
Manual operation:
See
CALCulate<n>:SGRam:FRAMe:SELect
<Frame> | <Time>
CALCulate<n>:SPECtrogram:FRAMe:SELect
<Frame> | <Time>
This command selects a specific frame for further analysis.
The command is available if no measurement is running or after a single sweep has
ended.
Parameters:
<Frame>
Selects a frame directly by the frame number. Valid if the time
stamp is off.
The range depends on the history depth.
Default unit: S
<Time>
Selects a frame via its time stamp. Valid if the time stamp is on.
The number is the distance to frame 0 in seconds. The range
depends on the history depth.
Example:
INIT:CONT OFF
Stop the continuous sweep.
CALC:SGR:FRAM:SEL -25
Selects frame number -25.
Manual operation:
See
CALCulate<n>:SGRam:HDEPth
<History>
CALCulate<n>:SPECtrogram:HDEPth
<History>
This command defines the number of frames to be stored in the R&S
FSV/A memory.
Analyzing Measurements (Basics)