Remote Commands
R&S
®
FSVA3000/ R&S
®
FSV3000
983
User Manual 1178.8520.02 ─ 01
CALCulate<n>:MARKer<m>[:STATe]
............................................................................... 987
................................................................................987
........................................................................................ 987
CALCulate<n>:DELTamarker<m>:AOFF
This command turns off
all
delta markers.
Suffix:
<n>
.
<m>
irrelevant
Example:
CALC:DELT:AOFF
Turns off all delta markers.
CALCulate<n>:DELTamarker<m>:LINK
<State>
This command links delta marker <m> to marker 1.
If you change the horizontal position (x-value) of marker 1, delta marker <m> changes
its horizontal position to the same value.
Tip
: to link any marker to a different marker than marker 1, use the
DELTamarker<ms>:LINK:TO:MARKer<md>
or
commands.
Suffix:
<n>
.
<m>
Parameters:
<State>
ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:DELT2:LINK ON
Manual operation:
See
CALCulate<n>:DELTamarker<ms>:LINK:TO:MARKer<md>
<State>
This command links delta marker <m1> to any active normal marker <m2>.
If you change the horizontal position of marker <m2>, delta marker <m1> changes its
horizontal position to the same value.
Suffix:
<n>
.
<ms>
source marker, see
Analyzing Measurements (Basics)