Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
189
User Manual 1173.9240.02 ─ 23
11.4.2.1
RF Input
INPut:ATTenuation:PROTection:RESet
............................................................................189
..........................................................................................190
.............................................................................................. 190
INPut:ATTenuation:PROTection:RESet
This command resets the attenuator and reconnects the RF input with the input mixer
after an overload condition occurred and the protection mechanism intervened. The
error status bit (bit 3 in the
STAT:QUES:POW
status register) and the
INPUT OVLD
message in the status bar are cleared.
(For details on the status register see the R&S
FSW User Manual).
The command works only if the overload condition has been eliminated first.
Usage:
Event
INPut:CONNector
<ConnType>
Determines whether the RF input data is taken from the RF input connector or the
optional Analog Baseband I connector. This command is only available if the Analog
Baseband interface (R&S
FSW-B71) is installed and active for input. It is not available
for the R&S
FSW67 or R&S
FSW85.
For more information on the Analog Baseband Interface (R&S
FSW-B71) see the
R&S
FSW I/Q Analyzer and I/Q Input User Manual.
Parameters:
<ConnType>
RF
RF input connector
AIQI
Analog Baseband I connector
*RST:
RF
Example:
INP:CONN:AIQI
Selects input from the analog baseband I connector.
Manual operation:
See
INPut:COUPling
<CouplingType>
This command selects the coupling type of the RF input.