Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
224
User Manual 1173.9240.02 ─ 23
Parameters:
<Coupling>
CENTer
Couples the frequency to the center frequency of the analyzer
MARKer1
Couples the frequency to the position of marker 1
OFF
Switches the frequency coupling off
*RST:
CENTer
Example:
PMET2:FREQ:LINK CENT
Couples the frequency to the center frequency of the analyzer
Manual operation:
See
[SENSe:]PMETer<p>:MTIMe
<Duration>
This command selects the duration of power sensor measurements.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<Duration>
SHORt | NORMal | LONG
*RST:
NORMal
Example:
PMET2:MTIM SHOR
Sets a short measurement duration for measurements of station-
ary high power signals for the selected power sensor.
Manual operation:
See
[SENSe:]PMETer<p>:MTIMe:AVERage:COUNt
<NumberReadings>
This command sets the number of power readings included in the averaging process of
power sensor measurements.
Extended averaging yields more stable results for power sensor measurements, espe-
cially for measurements on signals with a low power, because it minimizes the effects
of noise.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<NumberReadings> An average count of 0 or 1 performs one power reading.
Range:
0 to 256
Increment: binary steps (1, 2, 4, 8, ...)
Configuring the Measurement