Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
237
User Manual 1173.9240.02 ─ 23
This command is only available if external generator control is active (see
Example:
CORR:REC
Usage:
Event
Manual operation:
See
[SENSe:]CORRection[:STATe]
<State>
This command turns correction of measurement results (normalization) on and off.
The command is available after you have created a reference trace for the selected
measurement type with
[SENSe:]CORRection:COLLect[:ACQuire]
This command is only available if external generator control is active (see
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
CORR ON
Activates normalization.
Manual operation:
See
"Source Calibration Normalize"
[SENSe:]CORRection:TRANsducer:GENerate
<Name>
This command uses the normalized measurement data to generate a transducer factor
with up to 1001 points. The trace data is converted to a transducer with unit dB and
stored in a file with the specified name and the suffix
.trd
under
c:\r_s\instr\trd
. The frequency points are allocated in equidistant steps between
start and stop frequency.
The generated transducer factor can be further adapted using the commands descri-
bed in the "Remote Commands > Configuring the R&S
FSW > Working with Transduc-
ers" section in the R&S
FSW User Manual.
Parameters:
<Name>
'<name>
Example:
CORR:TRAN:GEN 'SMW200A1'
Creates the transducer file
C:\r_s\instr\trd\SMW200A.trd
.
Manual operation:
See
Programming Example for External Generator Control
The following example demonstrates how to work with an external generator in a
remote environment.
Configuring the Measurement