Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
254
User Manual 1173.9240.02 ─ 23
[SENSe:]ADEMod<n>:SPECtrum:BANDwidth[:RESolution]
<Bandwidth>
[SENSe:]ADEMod<n>:SPECtrum:BWIDth[:RESolution]
<Bandwidth>
Defines the resolution bandwidth for data acquisition.
From the specified RBW and the demodulation span set by
on page 254, the required measurement time is calculated. If the available measure-
ment time is not sufficient for the given bandwidth, the measurement time is set to its
maximum and the resolution bandwidth is increased to the resulting bandwidth.
This command is identical to
SENS:BAND:RES
, see the R&S
FSW User Manual.
Suffix:
<n>
.
irrelevant
Parameters:
<Bandwidth>
refer to data sheet
*RST:
61.2 kHz
Example:
ADEM:SPEC:BAND 61.2kHz
Sets the resolution bandwidth to 61.2 kHz.
[SENSe:]ADEMod<n>:SRATe?
This command returns the sample rate set up for the current analog demodulation
measurement.
Suffix:
<n>
.
irrelevant
Example:
ADEM:SRAT?
Returns the current sample rate.
Usage:
Query only
[SENSe:]BANDwidth:DEMod
<Bandwidth>
[SENSe:]BWIDth:DEMod
<Bandwidth>
This command sets the bandwidth for analog demodulation. Depending on the
selected demodulation bandwidth, the instrument selects the required sample rate.
For details on the correlation between demodulation bandwidth and sample rate refer
to
Chapter 4.3, "Sample Rate and Demodulation Bandwidth"
This command is identical to
SENS:ADEM:BAND:DEM
.
Parameters:
<Bandwidth>
*RST:
5 MHz
Example:
BAND:DEM 1MHz
Sets demodulation bandwidth to 1 MHz
Manual operation:
See
Configuring the Measurement