Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
263
User Manual 1173.9240.02 ─ 23
Parameters:
<Source>
IMMediate
Free Run
EXTernal
Trigger signal from the TRIGGER INPUT connector.
If the optional 2
GHz bandwidth extension (R&S
FSW-B2000) is
installed and active, this parameter activates the CH3 input con-
nector on the oscilloscope. Then the R&S
FSW triggers when
the signal fed into the CH3 input connector on the oscilloscope
meets or exceeds the specified trigger level.
Note:
In previous firmware versions, the external trigger was
connected to the CH2 input on the oscilloscope. As of firmware
version R&S
FSW 2.30, the
CH3
input on the oscilloscope must
be used!
EXT2
Trigger signal from the TRIGGER INPUT/OUTPUT connector.
Note: Connector must be configured for "Input".
EXT3
Trigger signal from the TRIGGER 3 INPUT/ OUTPUT connector.
Note: Connector must be configured for "Input".
RFPower
First intermediate frequency
Not available for input from the optional Digital Baseband Inter-
face or the optional Analog Baseband Interface.
IFPower
Second intermediate frequency
Not available for input from the optional Digital Baseband Inter-
face. For input from the optional Analog Baseband Interface, this
parameter is interpreted as
BBPower
for compatibility reasons.
IQPower
Magnitude of sampled I/Q data
For applications that process I/Q data, such as the I/Q Analyzer
or optional applications.
Not available for input from the optional Digital Baseband Inter-
face or the optional Analog Baseband Interface.
TIME
Time interval
BBPower
Baseband power (for digital input via the optional Digital Base-
band Interface
Baseband power (for digital input via the optional Digital Base-
band Interface or the optional Analog Baseband interface
PSEN
External power sensor
AF
AF power signal
FM
Configuring the Measurement