Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
278
User Manual 1173.9240.02 ─ 23
Parameters:
<Frequency>
numeric value
Range:
0 to 3 MHz
*RST:
15kHz
Example:
FILT:LPAS:FREQ:MAN 150kHz
The AF results are restricted to frequencies lower than 150 kHz.
Manual operation:
See
[SENSe:]FILTer<n>:LPASs:FREQuency:RELative
<FilterType>
This command selects the relative low pass filter type for the specified evaluation
For details on the low pass filter refer to
Suffix:
<n>
.
Parameters:
<FilterType>
5PCT | 10PCT | 25PCT
*RST:
25PCT
Example:
FILT:LPAS:FREQ:REL 25PCT
Selects the low pass filter as 25 % of the demodulation band-
width.
Manual operation:
See
[SENSe:]FILTer<n>:LPASs[:STATe]
<State>
This command activates/deactivates the selected low pass filter for the specified evalu-
ation.
For details on the low pass filter refer to
Suffix:
<n>
.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
FILT:LPAS ON
Activates the selected low pass filter.
Manual operation:
See
11.4.8.5
Defining the Scaling and Units
The scaling parameters define the range of the demodulated data to be displayed.
Configuring the Measurement