Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
282
User Manual 1173.9240.02 ─ 23
CONFigure:ADEMod:RESults:AM:DETector<det>:REFerence:MEAStoref
CONFigure:ADEMod:RESults:FM:DETector<det>:REFerence:MEAStoref
CONFigure:ADEMod:RESults:PM:DETector<det>:REFerence:MEAStoref
Sets the reference value to be used for relative demodulation results to the currently
measured value
for all relative detectors
.
If necessary, the detectors are activated.
A reference value
0
would provide infinite results and is thus automatically corrected to
0.1
.
Suffix:
<det>
.
irrelevant
Example:
See
CONFigure:ADEMod:RESults:PM:DETector<det>:
Usage:
Event
Manual operation:
See
CONFigure:ADEMod:RESults:AM:DETector<det>:MODE
<Mode>
CONFigure:ADEMod:RESults:FM:DETector<det>:MODE
<Mode>
CONFigure:ADEMod:RESults:PM:DETector<det>:MODE
<Mode>
Defines the mode with which the demodulation result is determined.
Suffix:
<det>
.
1: Positive peak; 2: Negative peak; 3: Average of positive and
negative peaks (+/-PK/2); 4: RMS
Detector function used for relative demodulation
Parameters:
<Mode>
WRITe
Overwrite mode: the detector value is overwritten by each
sweep. This is the default setting.
AVERage
The average result is determined over all sweeps.
MAXHold
The maximum value is determined over several sweeps and dis-
played. The R&S
FSW saves each result only if the new value is
greater than the previous one.
*RST:
WRITe
Example:
See
CONFigure:ADEMod:RESults:PM:DETector<det>:
Manual operation:
See
CONFigure:ADEMod:RESults:UNIT
<Unit>
This command selects the unit for relative demodulation results.
Configuring the Measurement