Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
330
User Manual 1173.9240.02 ─ 23
CALCulate<n>:MARKer<m>:SGRam:FRAMe
<Frame> | <Time>
CALCulate<n>:MARKer<m>:SPECtrogram:FRAMe
<Frame> | <Time>
This command positions a marker on a particular frame.
Suffix:
<n>
.
<m>
Parameters:
<Frame>
Selects a frame directly by the frame number. Valid if the time
stamp is off.
The range depends on the history depth.
<Time>
Selects a frame via its time stamp. Valid if the time stamp is on.
The number is the (negative) distance to frame 0 in seconds.
The range depends on the history depth.
Example:
CALC:MARK:SGR:FRAM -20
Sets the marker on the 20th frame before the present.
CALC:MARK2:SGR:FRAM -2s
Sets second marker on the frame 2 seconds ago.
CALCulate<n>:MARKer<m>:SGRam:SARea
<SearchArea>
CALCulate<n>:MARKer<m>:SPECtrogram:SARea
<SearchArea>
This command defines the marker search area for all spectrogram markers in the mea-
surement channel.
Suffix:
<n>, <m>
.
irrelevant
Parameters:
<SearchArea>
VISible
Performs a search within the visible frames.
Note that the command does not work if the spectrogram is not
visible for any reason (e.g. if the display update is off).
MEMory
Performs a search within all frames in the memory.
*RST:
VISible
CALCulate<n>:MARKer<m>:SGRam:XY:MAXimum[:PEAK]
CALCulate<n>:MARKer<m>:SPECtrogram:XY:MAXimum[:PEAK]
This command moves a marker to the highest level of the spectrogram.
Suffix:
<n>
.
<m>
Usage:
Event
Analyzing Results