Common Mode Offset
Sets the common mode offset. The setting is only available if a differential probe is
connected to the R&S
FSW.
If the probe is disconnected, the common mode offset of the probe is reset to 0.0
V.
For details, see the R&S
FSW I/Q Analyzer and I/Q Input User Manual.
Remote command:
[SENSe:]PROBe<p>:SETup:CMOFfset
Microbutton Action
Active Rohde
&
Schwarz probes (except for R&S
RT-ZS10E) have a configurable
microbutton on the probe head. By pressing this button, you can perform an action on
the instrument directly from the probe.
Select the action that you want to start from the probe:
"Run single"
Starts one data acquisition.
"No action"
Prevents unwanted actions due to unintended usage of the microbut-
ton.
Remote command:
5.3.1.6
External Generator Control Settings
Access
: INPUT/OUPUT > "External Generator Config"
The "External Generator" settings are available if the R&S
FSW External Generator
Control option is installed. For each measurement channel, you can configure one
external generator. To switch between different configurations, define multiple mea-
surement channels.