Configuration
R&S
®
FSW-K7
103
User Manual 1173.9240.02 ─ 23
"AC"
Connection through capacitor, removes unwanted DC and very low-
frequency components.
Remote command:
TRIGger[:SEQuence]:OSCilloscope:COUPling
Slope
For all trigger sources except time, you can define whether triggering occurs when the
signal rises to the trigger level or falls down to it.
When using the optional 2
GHz bandwidth extension (R&S
FSW-B2000) with an IF
power trigger, only rising slopes can be detected.
(For details see the R&S
FSW I/Q Analyzer and I/Q Input User Manual.)
Remote command:
Trigger Holdoff
Defines the minimum time (in seconds) that must pass between two trigger events.
Trigger events that occur during the holdoff time are ignored.
Remote command:
TRIGger[:SEQuence]:IFPower:HOLDoff
5.4.2 Trigger Input and Output Settings
Access
: "Overview" > "Trigger" > "Trigger In/Out" tab
.......................................................................................104
....................................................................................... 104
Trigger 2/3
Defines the usage of the variable TRIGGER INPUT/OUTPUT connectors, where:
"Trigger 2": TRIGGER INPUT/OUTPUT connector on the front panel
"Trigger 3": TRIGGER 3 INPUT/ OUTPUT connector on the rear panel
(Trigger 1 is INPUT only.)
Note:
Providing trigger signals as output is described in detail in the R&S
FSW User
Manual.
"Input"
The signal at the connector is used as an external trigger source by
the R&S
FSW. Trigger input parameters are available in the "Trigger"
dialog box.
"Output"
The R&S
FSW sends a trigger signal to the output connector to be
used by connected devices.
Further trigger parameters are available for the connector.
Note:
For offline AF or RF triggers, no output signal is provided.