Configuration
R&S
®
FSW-K7
130
User Manual 1173.9240.02 ─ 23
Adjusting settings automatically during triggered measurements
When you select an auto adjust function a measurement is performed to determine the
optimal settings. If you select an auto adjust function for a triggered measurement, you
are asked how the R&S
FSW should behave:
●
(default:) The measurement for adjustment waits for the next trigger
●
The measurement for adjustment is performed without waiting for a trigger.
The trigger source is temporarily set to "Free Run". After the measurement is com-
pleted, the original trigger source is restored. The trigger level is adjusted as fol-
lows:
–
For IF Power and RF Power triggers:
Trigger Level = Reference Level - 15 dB
–
For Video trigger:
Trigger Level = 85 %
Remote command:
[SENSe:]ADJust:CONFigure:TRIGger
Adjusting all Determinable Settings Automatically (Auto All)
...................................... 130
Adjusting the Center Frequency Automatically (Auto Freq)
........................................ 130
Setting the Reference Level Automatically (Auto Level)
............................................. 131
Resetting the Automatic Measurement Time (Meastime Auto)
...................................131
Changing the Automatic Measurement Time (Meastime Manual)
.............................. 131
............................................................................................... 131
............................................................................................... 132
Adjusting all Determinable Settings Automatically (Auto All)
Activates all automatic adjustment functions for the current measurement settings.
This includes:
●
●
●
Note:
MSRA / MSRT operating modes. In MSRA / MSRT operating mode, this function
is only available for the MSRA / MSRT Master, not the applications.
Remote command:
Adjusting the Center Frequency Automatically (Auto Freq)
The R&S
FSW adjusts the center frequency automatically.
The optimum center frequency is the frequency with the highest S/N ratio in the fre-
quency span. As this function uses the signal counter, it is intended for use with sinus-
oidal signals.
This function is not available for input from the optional Digital Baseband Interface.
Remote command:
Automatic Settings