Analysis
R&S
®
FSW-K7
143
User Manual 1173.9240.02 ─ 23
Remote command:
6.3.2 Marker Search Settings and Positioning Functions
Access
: "Overview" > "Analysis" > "Marker" > "Search"
or
: "Marker" > "Search"
Several functions are available to set the marker to a specific position very quickly and
easily, or to use the current marker position to define another characteristic value. In
order to determine the required marker position, searches may be performed. The
search results can be influenced by special settings.
The remote commands required to define these settings are described in
ter 11.8.1, "Working with Markers Remotely"
.........................................................................................143
............................................................................................144
6.3.2.1
Marker Search Settings
Access
: "Overview" > "Analysis" > "Marker" > "Search"
or
: "Marker" > "Search"
Markers are commonly used to determine peak values, i.e. maximum or minimum val-
ues, in the measured signal. Configuration settings allow you to influence the peak
search results.
........................................................................................ 143
Search Mode for Next Peak
Selects the search mode for the next peak search.
"Left"
Determines the next maximum/minimum to the left of the current
peak.
"Absolute"
Determines the next maximum/minimum to either side of the current
peak.
"Right"
Determines the next maximum/minimum to the right of the current
peak.
Remote command:
Chapter 11.8.1.5, "Positioning the Marker"
Peak Excursion
Defines the minimum level value by which a signal must rise or fall so that it is identi-
fied as a maximum or a minimum by the search functions.
Entries from 0
dB to 80
dB are allowed; the resolution is 0.1
dB. The default setting for
the peak excursion is 6
dB.
Working with Markers in the R&S FSW Analog Demodulation application