I/Q Data Import and Export
R&S
®
FSW-K7
163
User Manual 1173.9240.02 ─ 23
I/Q data can only be imported and exported in applications that process I/Q data, such
as the I/Q Analyzer or optional applications.
See the corresponding user manuals for those applications for details.
These functions are only available if no measurement is running.
In particular, if
is active, the import/export functions are
not available.
............................................................................ 163
........................................................................... 164
Import
Access
: "Save/Recall" > Import
Provides functions to import data.
I/Q Import ← Import
Opens a file selection dialog box to select an import file that contains I/Q data. This
function is only available in single sweep mode and only in applications that process
I/Q data, such as the I/Q Analyzer or optional applications.
Note that the I/Q data must have a specific format as described in the R&S
FSW I/Q
Analyzer and I/Q Input User Manual.
I/Q import is not available in MSRT mode.
Remote command:
Export
Access
: "Save/Recall" > Export
Opens a submenu to configure data export.
Export Trace to ASCII File ← Export
Opens a file selection dialog box and saves the selected trace in ASCII format (
.dat
)
to the specified file and directory.
The results are output in the same order as they are displayed on the screen: window
by window, trace by trace, and table row by table row.
Note:
Secure user mode.
In secure user mode, settings that are stored on the instrument are stored to volatile
memory, which is restricted to 256
MB. Thus, a "Memory full" error can occur although
the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
Import/Export Functions