I/Q Data Import and Export
R&S
®
FSW-K7
164
User Manual 1173.9240.02 ─ 23
For details, see "Protecting Data Using the Secure User Mode" in the "Data Manage-
ment" section of the R&S
FSW User Manual.
Remote command:
Trace Export Configuration ← Export
Opens the "Traces" dialog box to configure the trace and data export settings.
See
Chapter 6.2, "Trace / Data Export Configuration"
I/Q Export ← Export
Opens a file selection dialog box to define an export file name to which the I/Q data is
stored. This function is only available in single sweep mode.
Note:
Storing large amounts of I/Q data (several Gigabytes) can exceed the available
(internal) storage space on the R&S
FSW. In this case, it can be necessary to use an
external storage medium.
Note:
Secure user mode.
In secure user mode, settings that are stored on the instrument are stored to volatile
memory, which is restricted to 256
MB. Thus, a "Memory full" error can occur although
the hard disk indicates that storage space is still available.
To store data permanently, select an external storage location such as a USB memory
device.
For details, see "Protecting Data Using the Secure User Mode" in the "Data Manage-
ment" section of the R&S
FSW User Manual.
Remote command:
7.2 How to Export and Import I/Q Data
I/Q data can only be exported in applications that process I/Q data, such as the I/Q
Analyzer or optional applications.
Capturing and exporting I/Q data
1. Press the PRESET key.
2. Press the MODE key and select the R&S FSW Analog Demodulation application or
any other application that supports I/Q data.
3. Configure the data acquisition.
4. Press the RUN SINGLE key to perform a single sweep measurement.
5. Select the "Save" icon in the toolbar.
6. Select the "I/Q Export" softkey.
How to Export and Import I/Q Data