Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
196
User Manual 1173.9240.02 ─ 23
Table 11-3: Frequency ranges for pre-defined bands
Band
Frequency start [GHz]
Frequency stop [GHz]
KA (A) *)
26.5
40.0
Q
33.0
50.0
U
40.0
60.0
V
50.0
75.0
E
60.0
90.0
W
75.0
110.0
F
90.0
140.0
D
110.0
170.0
G
140.0
220.0
J
220.0
325.0
Y
325.0
500.0
USER
32.18
(default)
68.22
(default)
*)
The band formerly referred to as "A" is now named "KA".
[SENSe:]MIXer:HARMonic:HIGH:STATe
<State>
This command specifies whether a second (high) harmonic is to be used to cover the
band's frequency range.
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
MIX:HARM:HIGH:STAT ON
Manual operation:
See
[SENSe:]MIXer:HARMonic:HIGH[:VALue]
<HarmOrder>
This command specifies the harmonic order to be used for the high (second) range.
Parameters:
<HarmOrder
numeric value
Range:
2 to 61 (USER band); for other bands: see band
definition
Example:
MIX:HARM:HIGH 2
Manual operation:
See
Configuring the Measurement