Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
213
User Manual 1173.9240.02 ─ 23
INPut:DIQ:CDEVice
This command queries the current configuration and the status of the digital I/Q input
from the optional Digital Baseband Interface.
For details see the section "Interface Status Information" for the optional Digital Base-
band Interface in the R&S
FSW I/Q Analyzer User Manual.
Return values:
<ConnState>
Defines whether a device is connected or not.
0
No device is connected.
1
A device is connected.
<DeviceName>
Device ID of the connected device
<SerialNumber>
Serial number of the connected device
<PortName>
Port name used by the connected device
<SampleRate>
Maximum or currently used sample rate of the connected device
in Hz (depends on the used connection protocol version; indica-
ted by <SampleRateType> parameter)
<MaxTransferRate> Maximum data transfer rate of the connected device in Hz
<ConnProtState>
State of the connection protocol which is used to identify the
connected device.
Not Started
Has to be Started
Started
Passed
Failed
Done
<PRBSTestState>
State of the PRBS test.
Not Started
Has to be Started
Started
Passed
Failed
Done
<SampleRateType>
0
Maximum sample rate is displayed
1
Current sample rate is displayed
<FullScaleLevel>
The level (in dBm) that should correspond to an I/Q sample with
the magnitude "1" (if transferred from connected device);
If not available,
1.#QNAN
(not a number) is returned
Configuring the Measurement