Remote Commands for Analog Demodulation Measurements
R&S
®
FSW-K7
221
User Manual 1173.9240.02 ─ 23
Parameters:
ONCE
Example:
CAL:PMET2:ZERO:AUTO ONCE;*WAI
Starts zeroing the power sensor 2 and delays the execution of
further commands until zeroing is concluded.
Usage:
Event
Manual operation:
See
CALCulate<n>:PMETer<p>:RELative[:MAGNitude]
<RefValue>
This command defines the reference value for relative measurements.
Suffix:
<n>
.
<p>
1...4
Power sensor index
Parameters:
<RefValue>
Range:
-200 dBm to 200 dBm
*RST:
0
Example:
CALC:PMET2:REL -30
Sets the reference value for relative measurements to -30 dBm
for power sensor 2.
Manual operation:
See
CALCulate<n>:PMETer<p>:RELative[:MAGNitude]:AUTO ONCE
This command sets the current measurement result as the reference level for relative
measurements.
Suffix:
<n>
.
<p>
1...4
Power sensor index
Parameters:
ONCE
Example:
CALC:PMET2:REL:AUTO ONCE
Takes the current measurement value as reference value for rel-
ative measurements for power sensor 2.
Usage:
Event
Manual operation:
See
"Setting the Reference Level from the Measurement (Meas-
Configuring the Measurement